• DocumentCode
    806568
  • Title

    Computer-based waveform subtraction for measuring refraction of neuromuscular transmission

  • Author

    Young, S.T. ; Lin, H.J. ; Tsai, S.K.

  • Author_Institution
    Dept. of Surg., Veterans Gen. Hospital, Taiwan
  • Volume
    14
  • Issue
    4
  • fYear
    1995
  • Firstpage
    391
  • Lastpage
    394
  • Abstract
    The study demonstrated that computer-based waveform subtraction is a viable approach for the accurate determination of the refractory periods (RPs) of the neurally-evoked compound electromyograms (ncEMGs) and the refraction of neuromuscular transmission. Computer-based waveform subtraction can easily extract the R2 waveform from the complex ncEMGs and calculate accurate R2/R1 ratios. The method is more accurate and useful than the traditional “eyeballing,” which is the current method used. In the ncEMGs with twin stimuli, the interstimulus interval determines the time relationship between R2 and the R1. A short interstimulus interval obscures the R2 by the downward deflection of the R1. The emergence of the R2 is then delayed. Furthermore, the downward deflection elevates the baseline of the R2 waveform, making the measurement of the R2 amplitude difficult and unreliable by the traditional method. By use of computer-based waveform subtraction, the progression of R2 becomes smooth as the interstimulus intervals increase. The accurate R2 and the R2/R1 ratios are then available, and are suitable for evaluating the refraction of neuromuscular transmission. Applying the accurate R2/R1 in the study of neuromuscular physiology and pharmacology, a “dose-response” curve can be drawn and be used to investigate drug effects of many anesthetics and neuromuscular blocking drugs. In this study, the computer-based waveform subtraction resulted in an accurate dose response curve, similar to the traditional method but with a smaller refractory area
  • Keywords
    biomedical measurement; electromyography; medical signal processing; neurophysiology; R2 waveform; R2/R1 ratios; anesthetics; computer-based waveform subtraction; drug effects; interstimulus interval; neurally-evoked compound electromyograms; neuromuscular blocking drugs; neuromuscular physiology; neuromuscular transmission refraction measurement; pharmacology; Biomedical engineering; Biomedical measurements; Biomembranes; Cells (biology); Drugs; Economic indicators; Muscles; Neuromuscular; Surgery; Tin;
  • fLanguage
    English
  • Journal_Title
    Engineering in Medicine and Biology Magazine, IEEE
  • Publisher
    ieee
  • ISSN
    0739-5175
  • Type

    jour

  • DOI
    10.1109/51.395320
  • Filename
    395320