• DocumentCode
    808058
  • Title

    Evaluation of a Large Ge(Li) Detector for Low-Level Radionuclide Analysis

  • Author

    Brauer, F.P. ; Fager, J.E. ; Mitzlaff, W.A.

  • Author_Institution
    Battelle Pacific Northwest Laboratories Richland, Washington 99352
  • Volume
    20
  • Issue
    1
  • fYear
    1973
  • Firstpage
    57
  • Lastpage
    64
  • Abstract
    An evaluation of a 26.3% relative efficiency Ge(Li) detector for low-level radionuclide analysis has been made. The detector performance for sources mounted at the detector end cap was compared with the other smaller Ge(Li) detectors and with an opposed Ge(Li) two detector system. Factors studied included resolution, peak-to-Compton ratios, background, detection efficiency at the end cap, source geometry, and relative sensitivity. The 26.3% relative efficiency detector was found to give superior end cap efficiencies, peak-to-Compton ratios, and relative sensitivities without sacrificing performance in the other factors studied.
  • Keywords
    Assembly; Face detection; Gamma ray detection; Gamma ray detectors; Geometry; Laboratories; Manufacturing; Measurement standards; Molecular sieves; Spectroscopy;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1973.4326887
  • Filename
    4326887