Title :
A High Resolution Position Sensitive Detector for Ultraviolet and X-Ray Photons
Author :
Riegler, Guenter R. ; More, Keith A.
Author_Institution :
Space Physics Group Bendix Aerospace Systems Division Ann Arbor, Michigan 48107
Abstract :
A new type of position sensitive detection system for ultraviolet and x-ray photons and for particles is described. A compact laboratory model with one-dimensional readout was constructed and tested. The full width at half maximum response to single x-rays was typically 0.13 mm, but the position determination accuracy was comparable to the channel center-to-center spacing of 0.053 mm. The detection system can be readily adapted to two-dimensional readout. Potential applications of this device for single-photon and flux counting in ultraviolet and x-ray astronomy are discussed.
Keywords :
Diodes; Electrons; Energy resolution; Event detection; Glass; Laboratories; Optoelectronic and photonic sensors; Position sensitive particle detectors; X-ray detection; X-ray detectors;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1973.4326893