DocumentCode
808610
Title
Simultaneous measurement of thermal and ionization signals in a 60 g cryogenic germanium detector
Author
Barnes, P.D., Jr. ; Da Silva, A. ; Ellman, B. ; Emes, J. ; Haller, E.E. ; Lange, A.E. ; Ing, K. ; Ross, R.R. ; Sadoulet, B. ; Shutt, T. ; Smith, G. ; Stubbs, C. ; Wang, N. ; White, S. ; Yvon, D.
Author_Institution
Dept. of Phys., California Univ., Berkeley, CA, USA
Volume
39
Issue
5
fYear
1992
fDate
10/1/1992 12:00:00 AM
Firstpage
1237
Lastpage
1241
Abstract
The authors present results from a cryogenic detector instrumented to measure both the photon and ionization signals. The detector was a 60-g Ge disk operated at 30 mK. They have made significant improvements in the electronics noise, threshold, and energy resolution of the device. The baseline FWHM, corresponding to the total system noise, was 1.55 keV in the phonon channel and 1.59 keV in the ionization channel. At 60 keV the FWHM were 1.9 keV and 1.7 keV in the phonon and ionization channels, respectively. A discrimination between electron recoils and nuclear recoils at low energy was demonstrated
Keywords
cryogenics; elemental semiconductors; germanium; semiconductor counters; 60-g Ge disk; Ge; baseline FWHM; cryogenic detector; electron recoils; electronics noise; energy resolution; ionization channel; ionization signals; nuclear recoils; phonon channel; photon signals; thermal signals; threshold; total system noise; Cryogenics; Detectors; Electromagnetic scattering; Electrons; Energy resolution; Extraterrestrial measurements; Germanium; Ionization; Particle scattering; Phonons;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.173183
Filename
173183
Link To Document