• DocumentCode
    808610
  • Title

    Simultaneous measurement of thermal and ionization signals in a 60 g cryogenic germanium detector

  • Author

    Barnes, P.D., Jr. ; Da Silva, A. ; Ellman, B. ; Emes, J. ; Haller, E.E. ; Lange, A.E. ; Ing, K. ; Ross, R.R. ; Sadoulet, B. ; Shutt, T. ; Smith, G. ; Stubbs, C. ; Wang, N. ; White, S. ; Yvon, D.

  • Author_Institution
    Dept. of Phys., California Univ., Berkeley, CA, USA
  • Volume
    39
  • Issue
    5
  • fYear
    1992
  • fDate
    10/1/1992 12:00:00 AM
  • Firstpage
    1237
  • Lastpage
    1241
  • Abstract
    The authors present results from a cryogenic detector instrumented to measure both the photon and ionization signals. The detector was a 60-g Ge disk operated at 30 mK. They have made significant improvements in the electronics noise, threshold, and energy resolution of the device. The baseline FWHM, corresponding to the total system noise, was 1.55 keV in the phonon channel and 1.59 keV in the ionization channel. At 60 keV the FWHM were 1.9 keV and 1.7 keV in the phonon and ionization channels, respectively. A discrimination between electron recoils and nuclear recoils at low energy was demonstrated
  • Keywords
    cryogenics; elemental semiconductors; germanium; semiconductor counters; 60-g Ge disk; Ge; baseline FWHM; cryogenic detector; electron recoils; electronics noise; energy resolution; ionization channel; ionization signals; nuclear recoils; phonon channel; photon signals; thermal signals; threshold; total system noise; Cryogenics; Detectors; Electromagnetic scattering; Electrons; Energy resolution; Extraterrestrial measurements; Germanium; Ionization; Particle scattering; Phonons;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.173183
  • Filename
    173183