Title :
Operational Characteristics of Germanium Detectors at Higher Temperatures
Author :
Pehl, Richard H. ; Haller, Eugene E. ; Cordi, Richard C.
Author_Institution :
Lawrence Berkeley Laboratory University of California Berkeley, California 94720
Abstract :
Several important areas for the use of germanium spectrometers would be opened if these detectors could be successfully operated at temperatures considerably in excess of 77°K. Precise measurements on a series of detectors made from different high-purity germanium crystals and on one lithium-drifted germanium detector indicate that the resolution does not degrade significantly until nearly 150°K is reached. Similar-sized detectors, including the lithium-drifted device, exhibited essentially identical temperature-resolution relationships.
Keywords :
Crystals; Current measurement; Detectors; Diodes; Extraterrestrial measurements; Germanium; Leakage current; Steel; Temperature; Thermal resistance;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1973.4326953