• DocumentCode
    809103
  • Title

    High contrast, CdTe portal scanner for radiation therapy

  • Author

    Entine, G. ; Squillante, M.R. ; Hahn, R. ; Cirignano, L.J. ; McGann, W. ; Biggs, P.J.

  • Author_Institution
    Radiation Monitoring Devices Inc., Watertown, MA, USA
  • Volume
    39
  • Issue
    5
  • fYear
    1992
  • fDate
    10/1/1992 12:00:00 AM
  • Firstpage
    1480
  • Lastpage
    1484
  • Abstract
    A solid state, CdTe, linear array, portal imaging system is presented. One of the most promising new technologies for improving the quality of radiation therapy is the use of real-time systems to produce portal images. In the approach presented, a linear array of 256 CdTe photovoltaic detectors attached to a very compact linear scanner, all of which will be mounted in a cassette shaped package to be located under the patient table, is constructed. The high stopping power of the CdTe allows a high contrast image to be made using only a single linac pulse per line, resulting in a high contrast image in under 5 s
  • Keywords
    biomedical equipment; cadmium compounds; radiation therapy; semiconductor counters; CdTe; photovoltaic detectors; portal imaging system; radiation therapy; real-time systems; stopping power; Biomedical applications of radiation; Detectors; Linear particle accelerator; Packaging; Photovoltaic systems; Portals; Real time systems; Sensor arrays; Solar power generation; Solid state circuits;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.173228
  • Filename
    173228