Title :
Development of a 35-MHz piezo-composite ultrasound array for medical imaging
Author :
Cannat, Jonathan M. ; Williams, Jay A. ; Zhou, Qifa ; Ritter, Timothy A. ; Shung, K. Kirk
Author_Institution :
Dept. of Biomed. Eng., Southern California Univ., Los Angeles, CA, USA
Abstract :
This paper discusses the development of a 64-element 35-MHz composite ultrasonic array. This array was designed primarily for ocular imaging applications, and features 2-2 composite elements mechanically diced out of a fine-grain high-density Navy Type VI ceramic. Array elements were spaced at a 50-micron pitch, interconnected via a custom flexible circuit and matched to the 50-ohm system electronics via a 75-ohm transmission line coaxial cable. Elevation focusing was achieved using a cylindrically shaped epoxy lens. One functional 64-element array was fabricated and tested. Bandwidths averaging 55%, 23-dB insertion loss, and crosstalk less than -24 dB were measured. An image of a tungsten wire target phantom was acquired using a synthetic aperture reconstruction algorithm. The results from this imaging test demonstrate resolution exceeding 50 μm axially and 100 μm laterally.
Keywords :
biomedical ultrasonics; eye; image resolution; phantoms; tungsten; ultrasonic arrays; 23 dB; 35 MHz; 50 micron; 50 ohm; 75 ohm; elevation focusing; fine-grain high-density Navy Type VI ceramic; image resolution; medical imaging; ocular imaging; piezo-composite ultrasound array; synthetic aperture reconstruction algorithm; tungsten wire target phantom; Bioceramics; Biomedical imaging; Circuit testing; Coaxial cables; Distributed parameter circuits; Flexible printed circuits; Focusing; Integrated circuit interconnections; Lenses; Ultrasonic imaging; Ceramics; Computer Simulation; Computer-Aided Design; Electrochemistry; Equipment Design; Equipment Failure Analysis; Image Enhancement; Microelectrodes; Miniaturization; Models, Theoretical; Transducers; Ultrasonography;
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
DOI :
10.1109/TUFFC.2006.1588408