Title :
Reducing Gain Shifts in Photomultiplier Tubes
Author :
Cohn, Charles E.
Author_Institution :
Argonne National Laboratory Argonne, Illinois 60439
Abstract :
Gain shifts in photomultipliers from count-rate changes can be minimized by using only five active stages, making up the gain with a low-noise preamplifier. Further reductions sometimes result from a dim light shining on the dynode structure. Reductions to less than 1 part in 600 were seen.
Keywords :
Computer interfaces; Detectors; Electrodes; Laboratories; Photomultipliers; Preamplifiers; Pulse amplifiers; Resistors; Switches; Voltage;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :