• DocumentCode
    814130
  • Title

    Calibration of medical ultrasonic equipment-procedures and accuracy assessment

  • Author

    Preston, R.C. ; Bacon, D.R. ; Smith, Robert A.

  • Author_Institution
    NPL, Teddington, UK
  • Volume
    35
  • Issue
    2
  • fYear
    1988
  • fDate
    3/1/1988 12:00:00 AM
  • Firstpage
    110
  • Lastpage
    121
  • Abstract
    A beam-plotting facility has been developed to provide a reference measurement system for determining the acoustic output of medical ultrasonic equipment. It consists of two coordinate-positioning systems controlled by stepping motors and a minicomputer. One system is used for holding and manipulating an ultrasonic transducer and the other for a hydrophone. A membrane hydrophone made from polyvinylidene fluoride of 9- mu m thickness with an active element of 0.5-mm diameter is used for most measurements. The hydrophone is connected to an amplifier and digitizer, also controlled by the minicomputer, and the whole system has a measurement bandwidth of 75 MHz (-3 dB). A detailed description of this system is given together with a full assessment of measurement uncertainties and the methods used to correct for the effects of nonlinear distortion and spatial averaging. Typical overall uncertainties (95% confidence) for the determination of the peak-positive acoustic pressure, peak-negative acoustic pressure, spatial-peak pulse-average intensity and spatial-peak temporal-average intensity are +or-13%, +or-8%, +or-17%, and +or-23%, respectively.<>
  • Keywords
    biomedical ultrasonics; calibration; hydrophones; 75 MHz; PVDF; accuracy; acoustic output; acoustic pressure; amplifier; beam-plotting; calibration; coordinate-positioning systems; digitizer; medical ultrasonic equipment; membrane hydrophone; minicomputer; nonlinear distortion; reference measurement system; spatial averaging; stepping motors; ultrasonic transducer; Acoustic measurements; Acoustic pulses; Biomedical acoustics; Calibration; Control systems; Medical control systems; Microcomputers; Sonar equipment; Ultrasonic transducers; Ultrasonic variables measurement;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.4161
  • Filename
    4161