DocumentCode :
814296
Title :
On Increasing the Reliability of Fabrication and Handling of Large High Purity Germanium Detectors
Author :
Ewins, John ; Llacer, Jorge
Author_Institution :
Laboratory of Nuclear Medicine and Radiation Biology
Volume :
21
Issue :
1
fYear :
1974
Firstpage :
370
Lastpage :
373
Abstract :
A technique has been developed in the fabrication of high purity germanium gamma radiation detectors which consistently allows the application of large amounts of overvoltage. A chemically grown oxide layer to which Pd metallization adheres strongly forms a rugged surface barrier. Cycling temerature and between dry nitrogen and ambient air, changing contact fixtures, storage and transferring to different cryostats result in no significant changes in the initially good I-V and noise characteristics.
Keywords :
Etching; Fabrication; Fixtures; Gamma ray detectors; Germanium; Laboratories; Metallization; Methanol; Nitrogen; Nuclear medicine;