DocumentCode
814296
Title
On Increasing the Reliability of Fabrication and Handling of Large High Purity Germanium Detectors
Author
Ewins, John ; Llacer, Jorge
Author_Institution
Laboratory of Nuclear Medicine and Radiation Biology
Volume
21
Issue
1
fYear
1974
Firstpage
370
Lastpage
373
Abstract
A technique has been developed in the fabrication of high purity germanium gamma radiation detectors which consistently allows the application of large amounts of overvoltage. A chemically grown oxide layer to which Pd metallization adheres strongly forms a rugged surface barrier. Cycling temerature and between dry nitrogen and ambient air, changing contact fixtures, storage and transferring to different cryostats result in no significant changes in the initially good I-V and noise characteristics.
Keywords
Etching; Fabrication; Fixtures; Gamma ray detectors; Germanium; Laboratories; Metallization; Methanol; Nitrogen; Nuclear medicine;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1974.4327484
Filename
4327484
Link To Document