DocumentCode :
814760
Title :
Delay Line Readouts for High Purity Germanium Medical Imaging Cameras
Author :
Kaufman, Leon ; Camp, David C. ; McQuaid, James H. ; Armantrout, Guy A. ; Swierkowski, Steve P. ; Lee, Kai
Author_Institution :
University of California, San Francisco, Ca
Volume :
21
Issue :
1
fYear :
1974
Firstpage :
652
Lastpage :
657
Abstract :
High purity germanium offers excellent potential for use in nuclear medical imaging cameras. A position and energy readout technique using two inexpensive delay lines has been developed for these cameras. Results obtained with a 1-cm2 , 4 mm deep, 5×5 strip high purity germanium detector are 2.1 mm full width spatial resolution, a measured single strip resolution of 0.65 mm full width half maximum (FWHM), a 25 element uncollimated energy resolution of 2.95 keV FWHM, and 2.65 keV FWHM for a single central element at 140 keV.
Keywords :
Biomedical imaging; Cameras; Delay lines; Detectors; Energy resolution; Face detection; Germanium; Spatial resolution; Strips; Wires;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1974.4327528
Filename :
4327528
Link To Document :
بازگشت