DocumentCode
814837
Title
Chronic in Vivo Testing of a Fully Implantable Intracranial Pressure Sensor
Author
Meyer, Glenn A. ; Lyon, Warren C. ; Bustard, Thomas S.
Author_Institution
Department of Neurosurgery Medical College of Wisconsin 8700 West Wisconsin Avenue Milwaukee, Wisconsin
Volume
21
Issue
1
fYear
1974
Firstpage
702
Lastpage
706
Abstract
This paper describes the authors´ approach to the problem of long term intracranial pressure measurement. Several prior devices have not proven useful for clinical studies of more than a few days duration. Study of 22 of the authors´ devices in primates has established the validity of both engineering and medical design assumptions. Useful pressure data has been collected for up to 2 months. Redesign to decrease tambour permeability should allow a useful life of months or years.
Keywords
Batteries; Capacitors; Cranial pressure; In vivo; Intracranial pressure sensors; Scalp; Switches; Temperature; Testing; Transducers;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1974.4327536
Filename
4327536
Link To Document