• DocumentCode
    814837
  • Title

    Chronic in Vivo Testing of a Fully Implantable Intracranial Pressure Sensor

  • Author

    Meyer, Glenn A. ; Lyon, Warren C. ; Bustard, Thomas S.

  • Author_Institution
    Department of Neurosurgery Medical College of Wisconsin 8700 West Wisconsin Avenue Milwaukee, Wisconsin
  • Volume
    21
  • Issue
    1
  • fYear
    1974
  • Firstpage
    702
  • Lastpage
    706
  • Abstract
    This paper describes the authors´ approach to the problem of long term intracranial pressure measurement. Several prior devices have not proven useful for clinical studies of more than a few days duration. Study of 22 of the authors´ devices in primates has established the validity of both engineering and medical design assumptions. Useful pressure data has been collected for up to 2 months. Redesign to decrease tambour permeability should allow a useful life of months or years.
  • Keywords
    Batteries; Capacitors; Cranial pressure; In vivo; Intracranial pressure sensors; Scalp; Switches; Temperature; Testing; Transducers;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1974.4327536
  • Filename
    4327536