Title :
Sources of uncertainty in the nose-to-nose sampler calibration method
Author :
Paulter, N.G. ; Larson, D.R.
Author_Institution :
Dept. of Commerce, Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
Abstract :
We analyze the nose-to-nose (ntn) method for use as an accurate sampler calibration method. The variations in the measurement of the sampler impulse response using the ntn method are presented, and the validity of the assertion that the kick-out pulse is identical to the sampler impulse response is assessed. Temperature effects on the ntn method are also examined. Finally, the impact of uncertainties in the ntn calibration method on the uncertainties in reported pulse parameters is examined.
Keywords :
calibration; measurement uncertainty; transient response; impulse response; kick-out pulse; nose-to-nose sampler calibration method; pulse parameters; uncertainty; Bandwidth; Calibration; Oscilloscopes; Photoconductivity; Pulse generation; Pulse measurements; Sampling methods; Time domain analysis; Uncertainty; Voltage;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2003.817902