• DocumentCode
    814949
  • Title

    Sources of uncertainty in the nose-to-nose sampler calibration method

  • Author

    Paulter, N.G. ; Larson, D.R.

  • Author_Institution
    Dept. of Commerce, Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
  • Volume
    52
  • Issue
    5
  • fYear
    2003
  • Firstpage
    1618
  • Lastpage
    1626
  • Abstract
    We analyze the nose-to-nose (ntn) method for use as an accurate sampler calibration method. The variations in the measurement of the sampler impulse response using the ntn method are presented, and the validity of the assertion that the kick-out pulse is identical to the sampler impulse response is assessed. Temperature effects on the ntn method are also examined. Finally, the impact of uncertainties in the ntn calibration method on the uncertainties in reported pulse parameters is examined.
  • Keywords
    calibration; measurement uncertainty; transient response; impulse response; kick-out pulse; nose-to-nose sampler calibration method; pulse parameters; uncertainty; Bandwidth; Calibration; Oscilloscopes; Photoconductivity; Pulse generation; Pulse measurements; Sampling methods; Time domain analysis; Uncertainty; Voltage;