DocumentCode
815035
Title
Analysis of the effect of stray capacitance on an ac-based capacitance tomography transducer
Author
Yang, W.Q. ; Stott, A.L. ; Gamio, J.C.
Author_Institution
Dept. of Electr. Eng. & Electron., Univ. of Manchester Inst. of Sci. & Technol., UK
Volume
52
Issue
5
fYear
2003
Firstpage
1674
Lastpage
1681
Abstract
An ac-based capacitance transducer has been designed for use in electrical capacitance tomography (ECT) systems. The stray capacitance between the measurement electrodes and earth can be relatively large and may affect the performance of the transducer. This paper analyzes this effect, taking into account the ON-resistance of the CMOS switches in the circuit, the finite gain of op-amps and gain errors in the detection measuring unit, so that the overall measurement error can be estimated. It is shown that if the stray capacitance varies between 90 and 210 pF, the measurement error would be between +0.03% to -0.1%. This is negligibly small for ECT applications, confirming that this ac-based capacitance transducer is effectively stray-immune.
Keywords
capacitance; electric sensing devices; measurement errors; tomography; 90 to 210 pF; CMOS switches; ECT; ON-resistance; ac-based capacitance tomography transducer; electrical capacitance tomography; finite gain; gain errors; measurement electrodes; measurement error; overall measurement error; stray capacitance; Capacitance measurement; Earth; Electrical capacitance tomography; Electrodes; Gain measurement; Measurement errors; Operational amplifiers; Switches; Switching circuits; Transducers;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2003.817925
Filename
1240188
Link To Document