• DocumentCode
    815035
  • Title

    Analysis of the effect of stray capacitance on an ac-based capacitance tomography transducer

  • Author

    Yang, W.Q. ; Stott, A.L. ; Gamio, J.C.

  • Author_Institution
    Dept. of Electr. Eng. & Electron., Univ. of Manchester Inst. of Sci. & Technol., UK
  • Volume
    52
  • Issue
    5
  • fYear
    2003
  • Firstpage
    1674
  • Lastpage
    1681
  • Abstract
    An ac-based capacitance transducer has been designed for use in electrical capacitance tomography (ECT) systems. The stray capacitance between the measurement electrodes and earth can be relatively large and may affect the performance of the transducer. This paper analyzes this effect, taking into account the ON-resistance of the CMOS switches in the circuit, the finite gain of op-amps and gain errors in the detection measuring unit, so that the overall measurement error can be estimated. It is shown that if the stray capacitance varies between 90 and 210 pF, the measurement error would be between +0.03% to -0.1%. This is negligibly small for ECT applications, confirming that this ac-based capacitance transducer is effectively stray-immune.
  • Keywords
    capacitance; electric sensing devices; measurement errors; tomography; 90 to 210 pF; CMOS switches; ECT; ON-resistance; ac-based capacitance tomography transducer; electrical capacitance tomography; finite gain; gain errors; measurement electrodes; measurement error; overall measurement error; stray capacitance; Capacitance measurement; Earth; Electrical capacitance tomography; Electrodes; Gain measurement; Measurement errors; Operational amplifiers; Switches; Switching circuits; Transducers;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2003.817925
  • Filename
    1240188