DocumentCode
815038
Title
A Stable ADC for Use with Ge(Li) Detectors
Author
Lucas, C.H.
Author_Institution
Jet Propulsion Laboratory 4800 Oak Grove Drive Pasadena, CA 91103 (213) 354-6174
Volume
21
Issue
1
fYear
1974
Firstpage
798
Lastpage
804
Abstract
A new type of ADC is being developed to satisfy the requirements of the HEAO high spectral resolution gamma ray spectrometer. This 13 bit design utilizes a double successive approximation technique that overcomes the usual differential linearity problem. Stability of a few ppm/°C, differential linearity of 2% and integral linearity of 0.05% have been demonstrated.
Keywords
Detectors; Laboratories; Linearity; Propulsion; Pulse amplifiers; Pulse shaping methods; Resistors; Spectroscopy; Stability; Voltage;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1974.4327553
Filename
4327553