DocumentCode :
816309
Title :
CdTe Detectors in Radioimmunoassay Analysis
Author :
Zanio, K. ; Montano, H. ; Krajenbrink, F. ; Peterson, G.
Author_Institution :
Hughes Research Laboratories Malibu, California 90265
Volume :
22
Issue :
1
fYear :
1975
Firstpage :
422
Lastpage :
424
Abstract :
A well was constructed with cadmium telluride detectors and used at room temperature to detect 27.4 keV x-rays from 125I for the purpose of determining the concentration of carcinoembryonic antigen in human plasma. Concentrations were in good agreement with that earlier determined using a NaI well. For individual detectors approximately 1 to 2 cm2 in area and operating near 0°C, detector resolutions as low as 3 keV (FWHM) were obtained.
Keywords :
Apertures; Biomedical measurements; Humans; Laboratories; Plasma temperature; Plasma x-ray sources; Proteins; Testing; X-ray detection; X-ray detectors;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1975.4327673
Filename :
4327673
Link To Document :
بازگشت