DocumentCode :
81675
Title :
Performance Analysis for Steady Flow Generation and Improved Readout Signal in Amperometric Biosensors
Author :
Manfredonia, Ivano ; Barbieri, Matteo ; Verzicco, Roberto ; Stallo, Cosimo ; Ruggieri, Marina ; Lavecchia, Teresa ; Turemis, Mehmet ; Pezzotti, Gianni
Author_Institution :
Dept. of Electron. Eng., Univ. of Tor Vergata, Rome, Italy
Volume :
15
Issue :
9
fYear :
2015
fDate :
Sept. 2015
Firstpage :
5208
Lastpage :
5216
Abstract :
The use of peristaltic pumps as fluidic distribution microsystems in electrochemical biosensor devices operating with amperometric transducers can result in periodic spikes affecting the output signal. Mechanical components rotating inside the pumps are the main cause of this kind of disturbing noise. A novel fluidic setup based on simple fluid mechanics laws has been realized with no moving components when pump is on. It yields a fluctuation-free outcome signal that we compare with those produced by commercial peristaltic pumps. Evidence of the readout amperometric enhanced signal is shown thus confirming the successful implementation of more reliable biosensors and electrochemical signal transduction devices. Such stable and oscillation free amperometric signals are very desirable in view of biosensors integration in environmental monitoring platforms equipped with advanced communication technologies. An example is reported for their use in data mining systems equipped with a remote control for the real-time monitoring of pollutants. The cleanliness and the strength of the signal can adversely affect the reliability of the transmission of relevant information, especially when there are early warning civil purposes referred to a potential harmful contamination (i.e, water bodies pollution).
Keywords :
amperometric sensors; biosensors; data mining; flow measurement; flow sensors; pumps; readout electronics; reliability; transducers; amperometric biosensor; amperometric transducer; data mining system; early warning civil; electrochemical biosensor device; electrochemical signal transduction device; environmental monitoring platform; fluctuation-free outcome signal; fluidic distribution microsystem; fluidic setup; improved readout signal; mechanical component; oscillation free amperometric signal; peristaltic pump; potential harmful contamination; real-time pollutant monitoring; reliability; remote control; signal enhancement; simple fluid mechanics; steady flow generation; Biosensors; Electron tubes; Liquids; Pollution measurement; Pumps; Biosensors; Electrochemical devices; Fluidic Microsystems; Noise; Wireless Sensor Networks; electrochemical devices; fluidic microsystems; noise; wireless sensor networks;
fLanguage :
English
Journal_Title :
Sensors Journal, IEEE
Publisher :
ieee
ISSN :
1530-437X
Type :
jour
DOI :
10.1109/JSEN.2015.2439272
Filename :
7115018
Link To Document :