Title : 
Calibration of hydrophones based on reciprocity and time delay spectrometry
         
        
            Author : 
Ludwig, Gerhard ; Brendel, Klaus
         
        
            Author_Institution : 
Phys.-Tech. Bundesanstalt, Braunschweig, West Germany
         
        
        
        
        
            fDate : 
3/1/1988 12:00:00 AM
         
        
        
        
            Abstract : 
A brief survey is given of the calibration methods for hydrophones in the ultrasonic frequency range. The methods presently used in the Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, West Germany, for hydrophone calibrations in the frequency range from 1 to 15 MHz are the primary subject of concern. These methods are the two-transducer reciprocity method for the calibration at discrete frequencies, the time-delay-spectrometry substitution method for quasifrequency continuous calibrations, and the two-transducer reciprocity method with time-delay spectrometry, also for quasi-frequency continuous calibration. Compared with the calibration at discrete frequencies, the expenditure of time for a calibration is considerably reduced in the case of the last-mentioned method. The influencing parameters which affect the evaluation of the measurement uncertainty are briefly discussed for the calibration methods applied at the PTB.<>
         
        
            Keywords : 
calibration; hydrophones; ultrasonic transducers; 1 to 15 MHz; calibration; hydrophones; reciprocity; time delay spectrometry; two-transducer; ultrasonic; Biomembranes; Calibration; Delay effects; Frequency; Needles; Sonar equipment; Spectroscopy; Ultrasonic variables measurement; Voltage; Wavelength measurement;
         
        
        
            Journal_Title : 
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on