DocumentCode
818931
Title
A Multi-Wire Chamber System for Heavy Ion Beam Monitoring at the Bevalac
Author
Cuperus, J. ; Morgado, R.
Author_Institution
Lawrence Berkeley Laboratory University of California Berkeley, California
Volume
22
Issue
3
fYear
1975
fDate
6/1/1975 12:00:00 AM
Firstpage
1561
Lastpage
1564
Abstract
Horizontal and vertical integrated beam-current profiles are generated by a system of multi-wire chambers (32 wires/profile) operating in either the ionization or proportional mode. Sixteen distinct displays (1024 words) are digitally stored and any four may be simultaneously displayed. A new display can be generated at 64 ms intervals. A central control unit selects the mode of operation, the amount of delay after an appropriate trigger, the chamber integration time, and the particular chambers to be displayed. Operating in the proportional mode relativistic heavyion beam intensities as low as 104 charges cm-2 sec-1 can be detected.
Keywords
Assembly; Displays; Ion beams; Laboratories; Monitoring; Particle beams; Protons; Resonance; Structural beams; Wires;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1975.4327934