DocumentCode :
819106
Title :
Online dosimetry based on optically stimulated luminescence materials
Author :
Vaillé, J.R. ; Ravotti, F. ; Garcia, P. ; Glaser, M. ; Matias, S. ; Idri, K. ; Boch, J. ; Lorfèvre, E. ; McNulty, P.J. ; Saigné, F. ; Dusseau, L.
Author_Institution :
TS & PH Dept., CERN, Geneva, Switzerland
Volume :
52
Issue :
6
fYear :
2005
Firstpage :
2578
Lastpage :
2582
Abstract :
A version of the Optically Stimulated Luminescence (OSL) sensor specifically developed to monitor the dose online in radiation facilities is presented and calibrated with 60Co. The lowest dose measurable at the extremity of a 20 m cable is 0.3 mGy.
Keywords :
calibration; dosimetry; photodetectors; radiation monitoring; 0.3 mGy; 20 m; 60Co; calibration; online dose monitoring; online dosimetry; optically stimulated luminescence materials; optically stimulated luminescence sensor; radiation facilities; Dosimetry; Extraterrestrial measurements; Large Hadron Collider; Luminescence; Optical materials; Optical saturation; Optical sensors; Protons; Radiation monitoring; Stimulated emission; Online dosimetry; Optically Stimulated Luminescence (OSL); protons;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2005.860707
Filename :
1589241
Link To Document :
بازگشت