DocumentCode :
820203
Title :
A Pulse Processor for X-Ray Spectrometry with Si(Li)-Detectors
Author :
Kandiah, K. ; Smith, A.J. ; White, G.
Author_Institution :
United Kingdom Atomic Energy Authority Harwell, Didcot, Oxon, U.K.
Volume :
22
Issue :
5
fYear :
1975
Firstpage :
2058
Lastpage :
2065
Abstract :
This paper describes the general features of a pulse processor for use with opto-electronic charge sensitive head amplifiers particularly with Si(Li) X-ray detectors. The instrument contains time-variant filters with direct coupled amplifiers and comprehensive pile-up protection and live time correction. The test results indicate that the system can cope with extremely wide range of signal processing times, counting rates and changes of spectral content without any appreciable degradation in energy resolution and with useful output rates for input rates up to well over 50,000 events/sec.
Keywords :
Degradation; Energy resolution; Filters; Instruments; Protection; Pulse amplifiers; Signal processing; Spectroscopy; System testing; X-ray detectors;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1975.4328065
Filename :
4328065
Link To Document :
بازگشت