Title :
Investigation of coil geometry for transcutaneous energy transmission for artificial heart
Author :
Matsuki, H. ; Shiiki, M. ; Murakami, K. ; Yamamoto, T.
Author_Institution :
Dept. of Electr. Eng., Tohoku Univ., Sendai, Japan
fDate :
9/1/1992 12:00:00 AM
Abstract :
Transcutaneous energy transmission for a totally implantable artificial heart system is considered, and noncontact electric energy transmission by means of spiral coils with amorphous magnetic fiber is reported. Coil design for higher transmitting efficiency and a smaller temperature rise of the coil has been investigated. An experiment in vitro showed that temperature rise will be reduced within 3°C
Keywords :
artificial organs; coils; amorphous magnetic fiber; artificial heart; coil geometry; noncontact electric energy transmission; spiral coils; transcutaneous energy transmission; Artificial heart; Coils; Equations; Equivalent circuits; Geometry; Impedance; Iron; Resonance; Skin; Temperature;
Journal_Title :
Magnetics, IEEE Transactions on