• DocumentCode
    821226
  • Title

    Gamma-Ray Energy Deposition in Silicon Detectors

  • Author

    Lurie, N.A. ; Steinman, D.K. ; Harris, L., Jr. ; Wondra, J.P.

  • Author_Institution
    IRT Corporation, San Diego, California 92138
  • Volume
    22
  • Issue
    6
  • fYear
    1975
  • Firstpage
    2582
  • Lastpage
    2585
  • Abstract
    Calculations of the gamma-ray energy deposition spectrum have been performed for a silicon ionization dosimeter based on a silicon surface barrier detector. The technique used was a coupled photon-electron transport calculation using the three-dimensional Monte Carlo code, SANDYL. Comparison of calculated pulse-height spectra and measurements of standard calibrated gamma sources demonstrated the need to take into account the nonuniformity of the charge collection efficiency in the device. To this end, measurements were made of the radial dependence of the charge collection efficiency and sensitive volume. These results yielded an improved model and improved agreement between calcullation and experiment.
  • Keywords
    Charge measurement; Current measurement; Gamma ray detection; Gamma ray detectors; Ionization; Measurement standards; Monte Carlo methods; Pulse measurements; Silicon; Volume measurement;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1975.4328172
  • Filename
    4328172