Title :
Interlaboratory comparison of hydrophone calibrations
Author :
Preston, R.C. ; Bacon, D.R. ; Corbett, S.S. ; Harris, Gerald R. ; Lewin, Peter A. ; Mcgregor, James A. ; O´Brien, William D., Jr. ; Szabo, Thomas L.
Author_Institution :
NPL, Teddington, UK
fDate :
3/1/1988 12:00:00 AM
Abstract :
The results of an interlaboratory comparison of hydrophone calibration techniques in the frequency range 1-10 MHz are reported. Two membrane hydrophones were calculated to six laboratories, and each laboratory determined the end-of-cable loaded sensitivities using their normal calibration methods; these included optical interferometry, planar scanning, reciprocity combined with time-delay spectrometry, and suspended-sphere radiometry. After converting the results to end-of-cable open-circuit sensitivities, in most cases agreement between the various values was within +or-10% at all frequencies.<>
Keywords :
calibration; hydrophones; 1 to 10 MHz; end-of-cable loaded sensitivities; hydrophone calibrations; interlaboratory comparison; optical interferometry; planar scanning; reciprocity; suspended-sphere radiometry; time-delay spectrometry; Biomembranes; Calibration; Drugs; Frequency; Laboratories; Probes; Radio interferometry; Sonar equipment; Spectroscopy; Ultrasonic imaging;
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on