• DocumentCode
    821924
  • Title

    Triple-Coincidence with Automatic Chance Coincidence Correction

  • Author

    Chase, R.L.

  • Author_Institution
    Brookhaven National Laboratory Upton, New York 11973
  • Volume
    23
  • Issue
    1
  • fYear
    1976
  • Firstpage
    244
  • Lastpage
    246
  • Abstract
    The chance coincidences in a triple-coincidence circuit are of two types - partially correlated and entirely uncorrelated. Their relative importance depends on source strength and source and detector geometry so that the total chance correction cannot, in general, be calculated. The system to be described makes use of several delays and straightforward integrated circuit logic to provide independent evaluation of the two components of the chance coincidence rate.
  • Keywords
    Delay effects; Gamma rays; Geometry; Integrated circuit measurements; Laboratories; Logic circuits; Neutrons; Pulse measurements; Signal resolution; Solid scintillation detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1976.4328246
  • Filename
    4328246