    A Wide Range Counting-Campbelling Nuclear Instrumentation System

  • Author

    Oda, M. ; Wada, M. ; Badono, S.

  • Author_Institution
    Central Research Laboratory Mitsubishi Electric Corporation Minami-Shimizu, Amagasaki, Japan
  • Volume
    23
  • Issue
    1
  • fYear
    1976
  • Firstpage
    303
  • Lastpage
    306
  • Abstract
    Feature of the present system is the use of the developed low noise current-sensitive preamplifier, counting loss compensation pulse rate-meter and fast response true-mean-square circuit. Noise level of the preamplifier is 10 PA/¿Hz for 30 cable system. True mean square circuit operated at 100 KHz has a response time constant of 10 msec and a wide output range over 6 decades without level switching.