DocumentCode :
822608
Title :
Thrombus Detection Using I-125-Fibrinogen and a CdTe Probe
Author :
Garcia, Daniel A. ; Frisbie, James H. ; Tow, Donald E. ; Sasahara, Arthur A. ; Entine, Gerald
Author_Institution :
Veterans Administration Hospital West Roxbury, Massachusetts
Volume :
23
Issue :
1
fYear :
1976
Firstpage :
594
Lastpage :
598
Abstract :
A compact CdTe detector system was developed for use in a clinical screening test for venous thrombosis of the legs. Patients given intravenously administered autologous I-125-fibrinogen were probed externally at selected points on the thighs and calves for abnormal accumulations of radioactivity. Measurements made with the CdTe probe were compared to those obtained with a standard portable NaI detector system. The CdTe probe was the equal of the NaI detector in diagnostic capability. The compact design of the semiconductor system considerably eased the probing procedure, especially in bedridden patients with limited mobility of the extremities.
Keywords :
Blood; Detectors; Hospitals; Instruments; Leg; Life testing; Probes; Radiation monitoring; System testing; Thigh;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1976.4328312
Filename :
4328312
Link To Document :
بازگشت