• DocumentCode
    822769
  • Title

    A Detector System for In-Situ Spectrometric Analysis of 241Am and Pu in Soil

  • Author

    Kirby, J.A. ; Anspaugh, L.R. ; Phelps, P.L. ; Armantrout, G.A. ; Sawyer, D.

  • Author_Institution
    Division of Biomedical and Environmental Research University of California Lawrence Livermore Laboratory Livermore, California 94550
  • Volume
    23
  • Issue
    1
  • fYear
    1976
  • Firstpage
    683
  • Lastpage
    689
  • Abstract
    This system for quantitative analysis of Pu in soil via 241Am has four 2.5-mm high-purity Ge detectors of 33 cm2 total detecting surface area. These detectors are paralleled by gating circuitry to avoid the degradation of energy resolution associated with electronic output summing. In field tests the system was approximately three times as sensitive as a 70-cm3 Ge(Li) detector and approximately an order of magnitude more sensitive than the FIDLER system.
  • Keywords
    Capacitance; Circuit testing; Degradation; Energy resolution; Photonic crystals; Radiation detectors; Sensor arrays; Soil; Spectroscopy; System testing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1976.4328327
  • Filename
    4328327