Abstract
An ultra-high-sensitivity, versatile B -H looper/AC susceptometer is described. The design is based on the well-known inductance technique, and uses modern low-noise components to achieve the high sensitivity. The mechanical assembly is designed for use inside a glass dewar, enabling measurements to be made at any temperature between 4.2 K and 300 K. B -H loops at 5 kHz with up to 3.09 kOe applied field can be measured much faster and with greater sensitivity than with a SQUID (superconducting quantum interference device) magnetometer. Initial AC susceptibility can be measured over a frequency range of 12 Hz to 120 kHz
Keywords
inductance; magnetic hysteresis; magnetic susceptibility; magnetic variables measurement; 12 Hz to 120 kHz; 4.2 to 300 K; AC susceptibility; AC susceptometer; hysteresis looper; inductance technique; low-noise components; mechanical assembly; ultra-high-sensitivity; Assembly; Glass; Inductance; Interference; Magnetic field measurement; Mechanical variables measurement; Quantum mechanics; SQUIDs; Superconducting devices; Temperature sensors;