• DocumentCode
    825825
  • Title

    Characteristics of Large Intrinsic Germanium Detectors Operated at Elevated Temperatures

  • Author

    Nakano, G.H. ; Simpson, D.A. ; Imhof, W.L.

  • Author_Institution
    Lockheed Palo Alto Research Laboratory 3251 Hanover Street Palo Alto, California 94304
  • Volume
    24
  • Issue
    1
  • fYear
    1977
  • Firstpage
    68
  • Lastpage
    72
  • Abstract
    The operational characteristics of two large (25 and 75 cm3, coaxial) intrinsic germanium detectors have been investigated as a function of temperature in order to establish definitive base lines and design criteria for satellite-borne spectrometer systems. The selection of a cooling system and the optimization of the design are critically dependent upon the acceptable temperature limits for the operation of a germanium detector. To establish the usable range for operation we have studied the temperature dependence of critical parameters such as the energy resolution, pulse height and reverse current. The data, although based on only two detectors, indicate the possibility of operating large intrinsic detectors at temperatures significantly higher than that observed for Ge(Li) detectors of comparable size.
  • Keywords
    Coaxial components; Cooling; Design optimization; Detectors; Extraterrestrial measurements; Fingers; Germanium; Refrigerators; Spectroscopy; Temperature dependence;
  • fLanguage
    English