DocumentCode
826003
Title
Test & Evaluation of a Long Narrow Rectangular NaI(Tl) Polycrystalline Scintillator
Author
Hurley, J.Paul ; Moroney, John D., III ; Mastny, Gary F.
Author_Institution
Naval Undersea Center, San Diego, California 92132
Volume
24
Issue
1
fYear
1977
Firstpage
172
Lastpage
176
Abstract
A series of measurements was made to determine the performance of a long, rectangular, polycrystalline, NaI(Tl) scintillator. Using collimated sources of 137Cs and 22Na, scans were made along the scintillator to measure pulse height and energy resolution as a function of position. The measurements were made using first only one phototube at a time and then with both tubes. Measurements were also made with broad-beam, uncollimated radiation. The data were compiled and analyzed, and are presented in both graphical and tabular form.
Keywords
Collimators; Crystalline materials; Crystallization; Electric shock; Photomultipliers; Pulse amplifiers; Pulse measurements; Shape; Testing; Thermal resistance;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1977.4328664
Filename
4328664
Link To Document