DocumentCode :
827431
Title :
Positron Ring Cameras for Emission-Computed Tomography
Author :
Derenzo, Stephen E.
Author_Institution :
Donner Laboratory University of California Berkeley, California 94720
Volume :
24
Issue :
2
fYear :
1977
fDate :
4/1/1977 12:00:00 AM
Firstpage :
881
Lastpage :
885
Abstract :
A properly designed circular ring of NaI(Tl) crystals and suitable reconstruction algorithms can provide rapid, high-resolution transverse section images of positron-labeled compounds in the human body. The spatial resolution is limited by the crystal center-to-center spacing and by the positron range in tissue. Using 280 crystals with 10-mm center-to-center spacing and isotopes such as 68Ga. 11C, 18F, or 52Fe, a resolution of 7.5 mm full width at half maximum (FWHM) is possible. Improvements in this technology will involve larger numbers of smaller crystals, more efficient crystals (possibly bismuth germanate), or faster detectors (such as loaded organic scintillators).
Keywords :
Algorithm design and analysis; Bismuth; Cameras; Crystals; Humans; Isotopes; Positrons; Reconstruction algorithms; Spatial resolution; Tomography;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1977.4328802
Filename :
4328802
Link To Document :
بازگشت