• DocumentCode
    827523
  • Title

    Fast parametric beamformer for synthetic aperture imaging

  • Author

    Nikolov, Svetoslav Ivanov ; Jensen, Jørgen Arendt ; Tomov, Borislav Gueorguiev

  • Author_Institution
    Dept. of Electr. Eng., Tech. Univ. of Denmark, Lyngby
  • Volume
    55
  • Issue
    8
  • fYear
    2008
  • fDate
    8/1/2008 12:00:00 AM
  • Firstpage
    1755
  • Lastpage
    1767
  • Abstract
    This paper describes the design and implementation of a real-time delay-and-sum synthetic aperture beamformer. The beamforming delays and apodization coefficients are described parametrically. The image is viewed as a set of independent lines that are defined in 3D by their origin, direction, and inter-sample distance. The delay calculation is recursive and inspired by the coordinate rotation digital computer (CORDIC) algorithm. Only 3 parameters per channel and line are needed for their generation. The calculation of apodization coefficients is based on a piece- wise linear approximation. The implementation of the beamformer is optimized with respect to the architecture of a novel synthetic aperture real-time ultrasound scanner (SARUS), in which 4 channels are processed by the same set of field-programmable gate arrays (FPGA). In synthetic transmit aperture imaging, low-resolution images are formed after every emission. Summing all low-resolution images produces a perfectly focused high-resolution image. The design of the beamformer is modular, and a single beamformation unit can produce 4600 low-resolution images per second, each consisting of 32 lines and 1024 complex samples per line. In its present incarnation, 3 such modules fit in a single device. The summation of low-resolution images is performed internally in the FPGA to reduce the required bandwidth. The delays are calculated with a precision of 1/16th of a sample, and the apodization coefficients with 7-bit precision. The accumulation of low-resolution images is performed with 24-bit precision. The level of the side- and grating lobes, introduced by the use of integer numbers in the calculations and truncation of intermediate results, is below -86 dB from the peak.
  • Keywords
    biomedical ultrasonics; delays; field programmable gate arrays; image resolution; image scanners; medical image processing; piecewise linear techniques; CORDIC algorithm; FPGA; SARUS; apodization coefficients; beamforming delays; coordinate rotation digital computer algorithm; delay calculation; fast parametric beamformer; field-programmable gate arrays; inter-sample distance; low-resolution images; medical applications; perfectly focused high-resolution image; piece- wise linear approximation; real-time delay-and-sum synthetic aperture beamformer; single beamformation unit; synthetic aperture real-time ultrasound scanner; synthetic transmit aperture imaging; Array signal processing; Bandwidth; Computer architecture; Delay; Field programmable gate arrays; Focusing; Gratings; High-resolution imaging; Linear approximation; Ultrasonic imaging; Algorithms; Computer-Aided Design; Equipment Design; Equipment Failure Analysis; Image Enhancement; Image Interpretation, Computer-Assisted; Imaging, Three-Dimensional; Reproducibility of Results; Sensitivity and Specificity; Signal Processing, Computer-Assisted; Transducers; Ultrasonography;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2008.860
  • Filename
    4589189