DocumentCode
828732
Title
Ultra-low-power variable-resolution successive approximation ADC for biomedical application
Author
Le, H.P. ; Singh, J. ; Hiremath, L. ; Mallapur, V. ; Stojcevski, A.
Author_Institution
Centre of Telecommun. & Microelectron., Victoria Univ., Melbourne, Vic., Australia
Volume
41
Issue
11
fYear
2005
fDate
5/26/2005 12:00:00 AM
Firstpage
634
Lastpage
635
Abstract
An ultra-low-power variable-resolution successive approximation analogue-to-digital converter (ADC) is presented. A novel binary search algorithm architecture is proposed to replace the conventional digital-to-analogue converter to significantly reduce system area and power consumption. The proposed ADC consumes less than 22.2 μW of power at a conventional 2 V battery supply with a sampling rate of 200 samples/s, and standby power consumption of less than 1 μW.
Keywords
analogue-digital conversion; biomedical electronics; biomedical equipment; low-power electronics; search problems; 2 V; analogue-to-digital converter; binary search algorithm architecture; digital-to-analogue converter; power consumption;
fLanguage
English
Journal_Title
Electronics Letters
Publisher
iet
ISSN
0013-5194
Type
jour
DOI
10.1049/el:20050490
Filename
1437867
Link To Document