Title :
"Fungi" - A Radon Measuring Instrument with Fast Response
Author :
Negro, Vincent C. ; Watnick, Sidney
Author_Institution :
U. S. Department of Energy Environmental Measurements Laboratory New York, New York 10014
Abstract :
The FUNGI uses an electrostatic field to collect the daughter products of radon decay, and a solid state detector to resolve the energies of radium A (218po) and radium C´ (214po) alpha particles. This permits a great degree of flexibility in the measurement of radon and radon daughters. In addition, by limiting the instrument response to the radium A channel only, the response to changes in radon concentration of this instrument is ten times faster than other instruments, making it suitable for radon survey applications.
Keywords :
Detectors; Electrostatic measurements; Energy measurement; Equations; Filters; Fungi; Instruments; Pollution measurement; Sampling methods; Solid state circuits;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1978.4329405