DocumentCode
833618
Title
Avalanche Localization and Its Effects in Proportional Counters
Author
Fischer, J. ; Okuno, H. ; Walenta, A.H.
Author_Institution
Brookhaven National Laboratory Upton, New York 11973
Volume
25
Issue
1
fYear
1978
Firstpage
794
Lastpage
799
Abstract
Avalanche development around the anode wire in a gas proportional counter is investigated. In the region of proportional gas amplification, the avalanche is found to be well localized on one side of the anode wire, where the electrons arrive along the field lines from the point of primary ionization. Induced signals on electrodes surrounding the anode wire are used to measure the azimuthal position of the avalanche on the anode wire. Practical applications of the phenomena such as left-right assignment in drift chambers and measurement of the angular direction of the primary ionization electrons drifting towards the anode wire are discussed.
Keywords
Anodes; Collimators; Counting circuits; Electrodes; Electrons; Gain measurement; Ionization; Laboratories; Position measurement; Wire;