DocumentCode :
835844
Title :
A new heart-sounds gating device for medical imaging
Author :
Groch, Mark W. ; Domnanovich, James R. ; Erwin, William D.
Author_Institution :
Dept. of Med. Phys., Rush Graduate CoIl., Chicago, IL, USA
Volume :
39
Issue :
3
fYear :
1992
fDate :
3/1/1992 12:00:00 AM
Firstpage :
307
Lastpage :
310
Abstract :
A heart-sounds gating device has been designed and tested which identifies, individually, both the first (S1) and second (S2) heart sound from their timing relationship, providing two trigger points through the cardiac cycle for synchronizing medical images. The gate utilizes dynamically varying timing windows to anticipate the occurrence of S1 and S2. It has been initially applied to nuclear imaging of the cardiac bloodpool, but may be applied to any imaging modality requiring cardiac synchronization.
Keywords :
bioacoustics; biomedical equipment; cardiology; patient diagnosis; cardiac bloodpool; cardiac cycle; dynamically varying timing windows; heart-sounds gating device; medical images synchronization; medical imaging; nuclear imaging; timing relationship; trigger points; Biomedical imaging; Circuits; Electrocardiography; Filling; Heart; Logic; Medical diagnostic imaging; Nuclear medicine; Pulse amplifiers; Timing; Electrocardiography; Equipment Design; Gated Blood-Pool Imaging; Heart Rate; Humans; Phonocardiography; Reference Values; Stroke Volume;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.125016
Filename :
125016
Link To Document :
بازگشت