• DocumentCode
    835998
  • Title

    Improvement of the Light Collection in Scintillation Detectors

  • Author

    Klein, Horst ; Scholermann, Horst

  • Volume
    26
  • Issue
    1
  • fYear
    1979
  • Firstpage
    373
  • Lastpage
    377
  • Abstract
    The influence of geometrical and optical parameters on the light collection within scintillator assemblies has been investigated. Monte-Carlo calculations predicted a strong dependence on the design of the light guide and the effective reflectivity of its covering. Excellent agreement of experimental and theoretical data is shown for a small " standard"-scintillator (5.08 cm x 5.08 cm in diameter) as well as for largearea systems (5.08 cm x 25.4 cm in diameter). In both cases partially coated light guides should be used to restrict the local dependence of the light transmission to less than 2 % or 4 % resp. over the total volume. The averaged absolute transmission is calculated to be approximately 80 % or 45 % resp. for these optimized setups.
  • Keywords
    Assembly; Detectors; Energy resolution; Geometrical optics; Neutrons; Optical refraction; Photomultipliers; Production; Reflectivity; Scintillation counters;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1979.4329660
  • Filename
    4329660