DocumentCode
836222
Title
NEMA count-rate evaluation of the first and second generation of the Ecat Exact and Ecat Exact HR family of scanners
Author
Eriksson, Lars ; Wienhard, K. ; Eriksson, Marita ; Casey, M.E. ; Knoess, C. ; Bruckbauer, T. ; Hamill, J. ; Mulnix, T. ; Vollmar, S. ; Bendriem, B. ; Heiss, W.D. ; Nutt, R.
Author_Institution
CTI PET Syst., Inc., Knoxville, TN, USA
Volume
49
Issue
3
fYear
2002
fDate
6/1/2002 12:00:00 AM
Firstpage
640
Lastpage
643
Abstract
The first and second generation of the Exact and Exact HR family of scanners has been evaluated in terms of noise equivalent count rate (NEC) and count-rate capabilities. The new National Electrical Manufacturers Association standard was used for the evaluation. In spite of improved electronics and improved count-rate capabilities, the peak NEC was found to be fairly constant between the generations. The results are discussed in terms of the different electronic solutions for the two generations and its implications on system dead time and NEC count-rate capability.
Keywords
image scanners; positron emission tomography; Ecat Exact scanner; NEMA count-rate evaluation; National Electrical Manufacturers Association standard; count-rate capabilities; electronic solutions; first generation; noise equivalent count count-rate capability; noise equivalent count rate; positron camera systems; positron emission tomography count-rate performance; positron emission tomography imaging; second generation; system dead time; Cameras; Detectors; Helium; High-resolution imaging; Manufacturing; National electric code; Noise generators; Positron emission tomography; Production systems; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2002.1039542
Filename
1039542
Link To Document