• DocumentCode
    836222
  • Title

    NEMA count-rate evaluation of the first and second generation of the Ecat Exact and Ecat Exact HR family of scanners

  • Author

    Eriksson, Lars ; Wienhard, K. ; Eriksson, Marita ; Casey, M.E. ; Knoess, C. ; Bruckbauer, T. ; Hamill, J. ; Mulnix, T. ; Vollmar, S. ; Bendriem, B. ; Heiss, W.D. ; Nutt, R.

  • Author_Institution
    CTI PET Syst., Inc., Knoxville, TN, USA
  • Volume
    49
  • Issue
    3
  • fYear
    2002
  • fDate
    6/1/2002 12:00:00 AM
  • Firstpage
    640
  • Lastpage
    643
  • Abstract
    The first and second generation of the Exact and Exact HR family of scanners has been evaluated in terms of noise equivalent count rate (NEC) and count-rate capabilities. The new National Electrical Manufacturers Association standard was used for the evaluation. In spite of improved electronics and improved count-rate capabilities, the peak NEC was found to be fairly constant between the generations. The results are discussed in terms of the different electronic solutions for the two generations and its implications on system dead time and NEC count-rate capability.
  • Keywords
    image scanners; positron emission tomography; Ecat Exact scanner; NEMA count-rate evaluation; National Electrical Manufacturers Association standard; count-rate capabilities; electronic solutions; first generation; noise equivalent count count-rate capability; noise equivalent count rate; positron camera systems; positron emission tomography count-rate performance; positron emission tomography imaging; second generation; system dead time; Cameras; Detectors; Helium; High-resolution imaging; Manufacturing; National electric code; Noise generators; Positron emission tomography; Production systems; Spatial resolution;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2002.1039542
  • Filename
    1039542