DocumentCode
837156
Title
Measurement and calibration of a universal six-port network analyzer
Author
Lin, Weigan ; Ruan, Chengli
Author_Institution
Chengdu Inst. of Radio Eng., Sichuan, China
Volume
37
Issue
4
fYear
1989
fDate
4/1/1989 12:00:00 AM
Firstpage
734
Lastpage
742
Abstract
A general theory of measuring multiport networks is presented. In order to measure microwave multiport networks conveniently a method involving the stepwise reduction of the order of the network under test is suggested. All formulas for calibrating and measuring the triple six-port network analyzer, which is a universal six-port network analyzer, are given without any ambiguity. The procedures for calibration and measurement are very simple. No standard three-port network is needed to calibrate the six-port system. Finally, the error caused by nonideal isolation is discussed.<>
Keywords
S-matrix theory; calibration; error analysis; measurement errors; microwave circuits; microwave measurement; multiport networks; network analysers; calibration; error analysis; measurement; microwave multiport networks; universal six-port network analyzer; Calibration; Gold; Isolators; Microwave measurements; Microwave theory and techniques; Performance analysis; Performance evaluation; Read only memory; Scattering parameters; Testing;
fLanguage
English
Journal_Title
Microwave Theory and Techniques, IEEE Transactions on
Publisher
ieee
ISSN
0018-9480
Type
jour
DOI
10.1109/22.18847
Filename
18847
Link To Document