DocumentCode :
837381
Title :
Development of a flow-cell alpha detector utilizing microencapsulated CsI:Tl granules and silicon PIN-photodiodes
Author :
Tan, Hui ; DeVol, Timothy A.
Author_Institution :
Dept. of Environ. Eng. & Sci., Clemson Univ., SC, USA
Volume :
49
Issue :
3
fYear :
2002
fDate :
6/1/2002 12:00:00 AM
Firstpage :
1243
Lastpage :
1248
Abstract :
A flow-cell detector was developed utilizing microencapsulated granular CsI:Tl and silicon PIN-photodiodes for gross alpha radiation quantification. The CsI:Tl scintillator was crushed and sieved to 63-90 and 90-125 μm particle size and encapsulated with 10% (ratio of coating layer mass to particle mass) of parylene C polymer. The radiation detector consisted of a CsI:Tl flow-cell and two silicon PIN-photodiodes (PDs). Signals from the PDs were processed by charge-sensitive preamplifiers and shaping amplifiers, then digitized by a GaGe CompuScope 8012 A/PCI 12-bit dual-channel 50-MHz digital oscilloscope card residing in a Pentium III 450-MHz PC. The digitized signal was processed by a LabVIEW-based data-acquisition program for pulse-height and pulse-shape analyses in conjunction with a software-realized coincidence mode with a resolving time of 800 ns. By setting an appropriate region of interest (ROI) for alpha events in a dual parameter histogram, the spillover of beta events into alpha ROI was less than 3.9%. The initial detection efficiency for 233U was 8.8-19.5% and declined with time due to aqueous solution diffusing through the microencapsulant. The initial detection efficiency decreased with increased particle size, increased porosity, and increased contact time. The slow component of the detection efficiency "half-life" was measured to be 36 h when 1000 Bq mL-1 233U in 0.1 HNO3 solution circulated in a closed loop through the flow-cell at a flow rate of 1.0 ml min-1. The alpha minimum detectable concentration of a CsI:Tl flow-cell with approximate active volume of 0.075 ml was measured to be 12.4 Bq mL-1 for a 120-s count time.
Keywords :
alpha-particle detection; caesium compounds; high energy physics instrumentation computing; nuclear electronics; p-i-n photodiodes; preamplifiers; pulse height analysers; pulse shaping circuits; readout electronics; silicon radiation detectors; solid scintillation detectors; thallium; 120 s; 36 h; 450 MHz; 50 MHz; 63 to 90 micron; 90 to 125 micron; 233U; CsI:Tl; GaGe CompuScope 8012; HNO3; LabVIEW-based data-acquisition program; PIN-photodiodes; Si; alpha radiation quantification; charge-sensitive preamplifiers; coating layer mass; detection efficiency; digital oscilloscope card; digitized signal; dual parameter histogram; flow-cell alpha detector; microencapsulated granules; parylene C polymer; pulse-height analyses; pulse-shape analyses; scintillator; shaping amplifiers; Coatings; Oscilloscopes; Polymer films; Preamplifiers; Pulse amplifiers; Pulse shaping methods; Radiation detectors; Signal analysis; Signal processing; Silicon radiation detectors;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2002.1039645