• DocumentCode
    838911
  • Title

    Computed Tomographic Reconstruction of Beam Profiles with a Multi-Wire Chamber

  • Author

    Alonso, J.R. ; Tobias, C.A. ; Chu, W.T.

  • Author_Institution
    Lawrence Berkeley Laboratory
  • Volume
    26
  • Issue
    3
  • fYear
    1979
  • fDate
    6/1/1979 12:00:00 AM
  • Firstpage
    3077
  • Lastpage
    3079
  • Abstract
    MEDUSA (MEdical Dose Uniformity SAmpler), a 16 plane multi-wire proportional chamber, has been built to accurately measure beam profiles. The large number of planes allows for reconstruction of highly detailed beam intensity structures by means of Fourier convolution reconstruction techniques. This instrument is being used for verification and tuning of the Bevalac radiotherapy beams, but has potential applications in many beam profile monitoring situations.
  • Keywords
    Capacitors; Convolution; Image reconstruction; Instruments; Laboratories; Monitoring; Radio control; Tomography; Voltage; Wire;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1979.4329945
  • Filename