• DocumentCode
    840164
  • Title

    Capability evaluation of a sinogram error detection and correction method in computed tomography

  • Author

    Huang, Sung-Cheng ; Yu, Dan-Chu

  • Author_Institution
    Dept. of Radiol. Sci., Sch. of Med., California Univ., Los Angeles, CA, USA
  • Volume
    39
  • Issue
    4
  • fYear
    1992
  • fDate
    8/1/1992 12:00:00 AM
  • Firstpage
    1106
  • Lastpage
    1110
  • Abstract
    An iterative method for detecting and correcting errors in sinograms is described and its capability evaluated. The method is based on the use of a reconstruction-reprojection procedure to check the consistency of a sinogram. Adjustments of the sinogram are made iteratively to satisfy gradually the consistency requirement. The effectiveness of the method for detecting and correcting errors due to scanner detector problems is demonstrated with computer simulated data and with real data from position emission tomography. The potential usefulness of the method for quality control of scanners is discussed
  • Keywords
    computerised tomography; error correction; error detection; iterative methods; quality control; radioisotope scanning and imaging; capability; computed tomography; computer simulated data; consistency requirement; correction method; iterative method; position emission tomography; quality control; reconstruction-reprojection procedure; scanner detector problems; sinogram error detection; Computational modeling; Computed tomography; Computer errors; Computer simulation; Detectors; Error correction; Image reconstruction; Nuclear medicine; Positron emission tomography; Quality control;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.159768
  • Filename
    159768