DocumentCode :
841088
Title :
Analysis of a multielement ultrasound hyperthermia applicator
Author :
Benkeser, Paul J. ; Frizzell, Leon A. ; Goss, Stephen A. ; Cain, Charles A.
Author_Institution :
Sch. of Electr. Eng., Georgia Inst. of Technol., Atlanta, GA, USA
Volume :
36
Issue :
3
fYear :
1989
fDate :
5/1/1989 12:00:00 AM
Firstpage :
319
Lastpage :
325
Abstract :
An unfocused multielement ultrasound applicator was developed for hyperthermia treatment of superficial tumors. The applicator contains sixteen 3.8-cm/sup 2/ individually controllable elements on a 15.2-cm/sup 2/ piezoelectric ceramic plate. The acoustical power output of each element can be independently applied to facilitate uniform heating throughout the treatment area while minimizing undesired heating in normal tissues. The performance of the applicator was examined by measuring acoustical power output and beam profiles. The results of this analysis indicated that the applicator is capable of producing required therapeutic output levels with excellent localization and control of the power deposition.<>
Keywords :
acoustic arrays; biomedical equipment; biomedical ultrasonics; biothermics; radiation therapy; acoustic arrays; acoustical power output; beam profiles; biomedical ultrasonics; multielement ultrasound hyperthermia applicator; superficial tumors; Acoustic beams; Applicators; Frequency; Heat treatment; Hyperthermia; Neoplasms; Temperature control; Temperature distribution; Ultrasonic imaging; Ultrasonic variables measurement;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/58.19170
Filename :
19170
Link To Document :
بازگشت