    Electrical impedance tomography using the extended Kalman filter

  • Author

    Trigo, Flávio Celso ; Gonzalez-Lima, Raul ; Amato, Marcelo Brito Passos

  • Author_Institution
    Dept. of Mech. Eng., Univ. of Sao Paulo, Brazil
  • Volume
    51
  • Issue
    1
  • fYear
    2004
  • Firstpage
    72
  • Lastpage
    81
  • Abstract
    In this paper, we propose an algorithm that, using the extended Kalman filter, solves the inverse problem of estimating the conductivity/resistivity distribution in electrical impedance tomography (EIT). The algorithm estimates conductivity/resistivity in a wide range. The purpose of this investigation is to provide information for setting and controlling air volume and pressure delivered to patients under artificial ventilation. We show that, when the standard deviation of the measurement noise level raises up to 5% of the maximal measured voltage, the conductivity estimates converge to the expected vector within 7% accuracy of the maximal conductivity value, under numerical simulations, with spatial a priori information. A two-phase identification procedure is proposed. A cylindrical phantom with saline solution is used for experimental evaluation. An abrupt modification on the resistivity distribution of this solution is caused by the immersion of a glass object. Estimates of electrode contact impedances and images of the glass object are presented.
  • Keywords
    Kalman filters; bioelectric potentials; biomedical electrodes; electric impedance imaging; inverse problems; lung; parameter estimation; phantoms; physiological models; air volume control; artificial ventilation; conductivity distribution; cylindrical phantom; electrical impedance tomography; electrode contact impedances; extended Kalman filter; glass object images; glass object immersion; inverse problem; measurement noise level; pressure control; resistivity distribution; saline solution; standard deviation; two-phase identification procedure; Conductivity measurement; Glass; Impedance; Inverse problems; Measurement standards; Noise level; Noise measurement; Pressure control; Tomography; Ventilation; Algorithms; Atelectasis; Computer Simulation; Diagnosis, Computer-Assisted; Humans; Lung; Models, Biological; Phantoms, Imaging; Plethysmography, Impedance; Reproducibility of Results; Sensitivity and Specificity; Signal Processing, Computer-Assisted; Thorax;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2003.820389
  • Filename
    1253996