• DocumentCode
    845059
  • Title

    Characteristics of a radiation-charged electret dosimeter

  • Author

    Fallone, B.G. ; MacDonald, B.A. ; Ryner, L.R.

  • Author_Institution
    Dept. of Radiat. Oncology, McGill Univ., Montreal, Que., Canada
  • Volume
    28
  • Issue
    1
  • fYear
    1993
  • fDate
    2/1/1993 12:00:00 AM
  • Firstpage
    143
  • Lastpage
    148
  • Abstract
    The radiation-charged electret ionization chamber uses an electret formed in situ by applying a bias voltage to the chamber electrodes while the chamber is being irradiated with X-rays. An electrometer permits measurement of the charge on the electret surface before and after the use of the chamber as a dosimeter. The authors present measurements of dosimeter range, sensitivity and energy dependence as functions of bias potential, electrode separation, guard ring size and various materials. Reproducibility permits detection as low as 80 μGy, with range to ~5 Gy, depending on the design of the chamber, with charge retention properties similar to electrets formed isothermally from corona discharge
  • Keywords
    charge measurement; dosimeters; electrets; ionisation chambers; 8×10-5 to 5 Gy; X-ray irradiation; bias potential; bias voltage; charge retention; corona discharge; dosimeter range; electrode separation; electrometer; energy dependence; guard ring size; ionization chamber; radiation-charged electret dosimeter; sensitivity; Charge measurement; Current measurement; Electrets; Electrodes; Energy measurement; Ionization chambers; Ionizing radiation; Size measurement; Voltage; X-rays;
  • fLanguage
    English
  • Journal_Title
    Electrical Insulation, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9367
  • Type

    jour

  • DOI
    10.1109/14.192251
  • Filename
    192251