DocumentCode :
845310
Title :
Dose Reduction in X-Ray Computed Tomography Using a Generalized Filter
Author :
Reed, I.S. ; Glenn, W.Y. ; Chang, C.M. ; Truong, T.K. ; Kwoh, Y.S.
Volume :
26
Issue :
2
fYear :
1979
fDate :
4/1/1979 12:00:00 AM
Firstpage :
2904
Lastpage :
2909
Abstract :
Previously with the use of computerized phantoms and simulation, it was conjectured that the x-ray dosage could be reduced by using the appropriate generalized |¿| - filter for x-ray reconstruction. In this paper, this conjecture has been confirmed by experiments that use data fromanEMI 5005 CT-scanner. Explicitly it is shown that the x-ray dosage can be lowered to 1/11 of the standard dosage while still maintaining a reconstruction accuracy which is acceptable for some applications.
Keywords :
Back; Computational modeling; Computed tomography; Convolution; Cutoff frequency; Filters; Hospitals; Imaging phantoms; Interpolation; X-ray imaging;