Title :
Properties of a one-cylinder-type standard cell
Author :
Shimazaki, Kazumi ; Shida, Katstjnori
Author_Institution :
Electrotech. Lab., Tsukuba, Japan
fDate :
4/1/1989 12:00:00 AM
Abstract :
Two kinds of one-cylinder-type standard cells are described. Compared to a normal H-shaped standard cell, they have the advantage of improved null-detector resolution for electromagnetic force (EMF) measurements of standard cells because a low internal resistance is easily obtained in a compact volume. In a poorly temperature-regulated enclosure, the one-cylinder-type standard cells also produce a stable EMF measurement because the two electrodes of the standard cell are close to each other
Keywords :
measurement standards; voltage measurement; EM force measurement; low internal resistance; null-detector resolution; one-cylinder-type standard cell; poorly temperature-regulated enclosure; stable EMF measurement; voltage standard; Detectors; Electrical resistance measurement; Electrodes; Laboratories; Legged locomotion; Measurement standards; Q measurement; Stability; Standards development; Voltage measurement;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on