DocumentCode :
845582
Title :
High accuracy calculable AC-DC transfer standards for the LF-30 MHz frequency range
Author :
Nomair, M. ; Harmans, Kees J.P.M.
Author_Institution :
Van Swinden Lab., Delft, Netherlands
Volume :
38
Issue :
2
fYear :
1989
fDate :
4/1/1989 12:00:00 AM
Firstpage :
342
Lastpage :
345
Abstract :
A model for a calculable AC-DC thermal transfer device is described. It is based on the transmission-line theory, augmented by corrections for skin effect and connection admittance. Experimental verification on four devices shows good to excellent agreement between the calculated and measured results. This enables the establishment of AC-DC transfer standards with total uncertainties ranging from less than 5 p.p.m. at 1 MHz to 100 p.p.m. at 30 MHz. Potential for extension to higher frequencies is discussed
Keywords :
electric current measurement; measurement standards; skin effect; transmission line theory; voltage measurement; 1 to 30 MHz; AC-DC transfer standards; connection admittance; model; skin effect; thermal transfer; transmission-line theory; Admittance; Connectors; Frequency; Impedance; Laboratories; Resistors; Tellurium; Transmission line theory; Voltage; Wire;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.192303
Filename :
192303
Link To Document :
بازگشت