Abstract :
Waveform sampling is an appealing technique for instruments requiring precision time and pulse-height measurements. Sampling each photomultiplier tube (PMT) waveform at oscilloscope-like rates of several gigasamples per second enables one to process PMT signals digitally, which in turn makes it straightforward to optimize timing resolution and amplitude (energy and position) resolution in response to calibration effects, pile-up effects, and other systematic sources of waveform variation. We describe a system design and preliminary implementation that neatly maps waveform-sampling technology onto the LaPET prototype whole-body time-of-flight PET scanner that serves as the platform for testing this new technology.
Keywords :
biomedical electronics; calibration; photomultipliers; positron emission tomography; sampling methods; waveform analysis; LaPET prototype; PMT; calibration effects; photomultiplier tube; waveform-sampling electronics; whole-body time-of-flight PET scanner; Arrays; Crystals; Detectors; Energy resolution; Field programmable gate arrays; Positron emission tomography; Timing; Fast timing; PET instrumentation; field-programmable gate array (FPGA); front-end electronics; time-of-flight PET; trigger systems; waveform sampling;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2014.2303119
Filename :
6800138
Link To Document :
بازگشت